What it is

A steam sterilizer, often called an autoclave, is a machine that sterilizes loads with saturated steam under pressure in a sealed chamber. Hospitals use it for reusable medical devices. EN 285 covers large steam sterilizers and EN 13060 covers small ones. The process itself is validated separately.

Key Takeaways

  • Steam sterilizer is the standards term. Autoclave is the common name.
  • EN 285 covers large machines and EN 13060 covers small ones.
  • Air removal matters as much as temperature for wrapped and porous loads.
  • The machine standard does not validate the process. That is done on site.

Why it matters

A steam sterilizer exposes a load to saturated steam at a set temperature for a set holding time inside a pressure chamber. Hospital sterile services departments use it for reusable medical devices such as wrapped surgical instruments. Steam must reach every surface, so air has to be removed first. Sterilizers for wrapped and porous loads do this with a vacuum system, and usually have a heated jacket and an air detector. European standards split the machines by size. EN 285 covers large steam sterilizers: those that hold at least one sterilization module or have a chamber of 60 litres or more. EN 13060 covers small steam sterilizers below that size and sorts their cycles into types B, N and S by the loads they can process. Both are machine standards. The process run in the machine is validated separately, usually to ISO 17665. The chamber is pressure equipment, and a sterilizer for medical devices is regulated as a medical device itself. A hospital tender for a steam sterilizer states the chamber size, doors, cycles, steam source, records, tests and documents it wants. Use the edition of each standard that the tender names.

How OdysseyGPT uses it

Document Review checks a hospital's tender package for a steam sterilizer against your datasheets, standard specifications and test procedures, and drafts a compliance matrix and deviation list with a quote for each row. It reads documents only. It does not test a sterilizer, validate a process or read drawings, and your engineers set every status.

Evaluation questions

What does a hospital tender for a steam sterilizer specify?

Usually chamber size, single or double doors, the cycles and temperatures wanted, cycle time, the steam source, utilities, records and data interfaces, the tests to be witnessed, and the validation and handover documents.

Frequently asked questions

What is the difference between a steam sterilizer and an autoclave?

In hospital use they are the same machine. Steam sterilizer is the term the standards use. Autoclave is the everyday word, and it is also used for pressure vessels in other industries that have nothing to do with sterilization.

What is the difference between a large and a small steam sterilizer?

Size, and the standard that applies. EN 285 covers large sterilizers, which hold at least one sterilization module or have a chamber of 60 litres or more. EN 13060 covers small sterilizers below that, such as many benchtop machines.

What is a porous-load sterilizer?

A steam sterilizer built for wrapped goods and porous items such as textiles, which trap air and moisture. It has a vacuum system to remove air before steam is admitted and to dry the load at the end of the cycle.

References

Health Technical Memorandum 01-01 Part C: Steam sterilization

Department of Health / NHS England

Free guidance for England.

Visit source

Small steam sterilizers and steam boilers: new standards for safer sterilization

CEN and CENELEC

Visit source

EN 285: Sterilization. Steam sterilizers. Large sterilizers

CEN

Paid standard.

Related Pages